• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUADRA ASSURA MP ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUADRA ASSURA MP ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3369-40Q
Device Problems Misconnection (1399); Fitting Problem (2183)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/06/2020
Event Type  malfunction  
Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
It was reported that the patient presented for an implant procedure.During the procedure, there was difficulty to seat the left ventricular lead fully with the set screw in the header.While attempting to confirm that the lead was not stuck in the header, the gasket was removed to loosen the set screw as the wrench would not turn the screw.The device was replaced.The patient was in stable condition.
 
Manufacturer Narrative
The reported event of header anomaly could not be confirmed in the lab.Interrogation of the device revealed the device was above the elective replacement indicator when received.The device was tested on the bench, and no anomalies were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA MP ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
MDR Report Key9753746
MDR Text Key181152469
Report Number2017865-2020-02610
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734508377
UDI-Public05414734508377
Combination Product (y/n)N
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 06/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/26/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2021
Device Model NumberCD3369-40Q
Device Catalogue NumberCD3369-40Q
Device Lot NumberA000091556
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/01/2020
Was the Report Sent to FDA? No
Date Manufacturer Received06/03/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age78 YR
Patient Weight73
-
-