The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history of the reported lot did not indicate a lot specific quality issue.The investigation determined that the reported failure to advance resulting in steerable guide catheter (sgc) soft tip damage appeared to be related to patient anatomy.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|
This is being filed to report soft tip tear.It was reported that this was a mitraclip procedure to treat mixed mitral regurgitation (mr) with a grade of 4.During insertion, the steerable guide catheter (sgc) met resistance with the anatomy and did not reach the septum.The sgc was removed and tears on the soft tip were observed.It was noted that the course of the vein was altered by cancer.The physician stated that stenosis developed in the veins caused by cancer which affected the right femoral vein access.Therefore, the second sgc was advanced through the left groin.Two clips were implanted, reducing mr to 2.No additional information was provided.
|