BOSTON SCIENTIFIC CORPORATION COMET; TRANSDUCER, PRESSURE, CATHETER TIP Back to Search Results
Model Number 8900
Device Problem Failure to Zero (1683)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/12/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Device returned to manufacturer.The comet was returned and analysis was completed.The tip, device shaft, sensor port and the coefficient values were examined for damage or any irregularities.The devices shaft showed two kinks located 176.5cm and at the occ handle.There was some peeled coating at the 176.5 location.The occ handle was connected to the ffr link to verify the signal strength.The signal was present as designed.The sensor port showed no residue of body fluids.The occ handle was then connected to the bench top testing equipment and the wire was inserted into the pressure chamber.The pressure was increased to verify the sensor was indeed reacting to the pressure increases and decreases.The pressure sensor functioned as designed.The coefficient was confirmed to be within specification the occ handle was again connected to the ffr link.The device was then connected to the polaris (ilab) test equipment via bluetooth signal.The wire communicated to the polaris system and zeroed as designed.The wire was inserted into the test pressure chamber and the wire transferred a pressure waveform to the polaris which indicates a functioning wire.The wire was removed from the occ handle with no issues.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.
 
Event Description
Reportable based on analysis completed 4 feb 2020.It was reported that recognition issue occurred.It was reported that the comet pressure guidewire was not recognized.The procedure was completed with another of the same device.There were no patient complications reported.However, returned device analysis revealed peeled coating.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMET
Type of Device
TRANSDUCER, PRESSURE, CATHETER TIP
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key9758208
MDR Text Key181189371
Report Number2134265-2020-02102
Device Sequence Number1
Product Code DXO
UDI-Device Identifier08714729904403
UDI-Public08714729904403
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K151610
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 02/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/15/2021
Device Model Number8900
Device Catalogue Number8900
Device Lot Number0023653441
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/22/2020
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/04/2020
Initial Date FDA Received02/26/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/16/2019
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-