• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DENTSPLY VDW GMBH VDW SILVER RECIPROCATING MOTOR; CONTROLLER, FOOT, HANDPIECE AND CORD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTSPLY VDW GMBH VDW SILVER RECIPROCATING MOTOR; CONTROLLER, FOOT, HANDPIECE AND CORD Back to Search Results
Catalog Number V041163000000
Device Problem Retraction Problem (1536)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
There has been a previous report received where stalling has caused file separation.Since separation of a file could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803 please note that while this product is not sold in the us, it is considered similar to products that are marketed in the us by dentsply (b)(4).Battery (voltage breaks down under load) defective, replaced.Micromotor (bent shaft) defective, replaced.Sent in without foot control and power supply.Function test and test measurements without errors.Test carried out according to (b)(4).
 
Event Description
In this event it was reported that a silver reciproc motor turns unbalanced and sometimes completely stops.No patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VDW SILVER RECIPROCATING MOTOR
Type of Device
CONTROLLER, FOOT, HANDPIECE AND CORD
Manufacturer (Section D)
DENTSPLY VDW GMBH
bayerwaldstrasse 15
munich, 81737
GM  81737
Manufacturer (Section G)
ADVANCED TECHNOLOGY RESEARCH (A.T.R.) S.R.L.
via san donato 1
pistoia, 51100
IT   51100
Manufacturer Contact
karl nittinger
221 w. philadelphia st.
suite 60w
york, pa 
8494424
MDR Report Key9764281
MDR Text Key181635760
Report Number9611053-2020-00041
Device Sequence Number1
Product Code EBW
Combination Product (y/n)N
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial
Report Date 02/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberV041163000000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/20/2020
Initial Date Manufacturer Received 01/28/2020
Initial Date FDA Received02/27/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-