CERNER CORPORATION CERNER'S MILLENNIUM POWERCHART®, FIRSTNET®, SURGICAL POWERCHART® AND DISCERN®; SOFTWARE Back to Search Results
Model Number 2007.19 THROUGH 2018.09
Device Problems Computer Software Problem (1112); Patient Data Problem (3197)
Patient Problems Patient Problem/Medical Problem (2688); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/18/2019
Event Type  malfunction  
Manufacturer Narrative
Cerner distributed a flash notification on february 24, 2020 to all potentially impacted client sites.The software notification includes a description of the issue and a software modification is being developed to address the issue for all sites that could be potentially impacted.Cerner corporation will provide a follow-up report when the software modification is available.
 
Event Description
The software product mentioned in this medwatch report may not be, by definition, a medical device; however, cerner has chosen to file this medwatch report in order to voluntarily notify the fda of a malfunction associated with this software product.The company's filing of this medwatch report does not signify cerner's belief or understanding that medical device reports are required to be filed for products such as cerner's clinical notes®, nor are these products currently actively regulated by the fda.This report documents information related to an issue identified with functionality included in cerner's millennium clinical notes®.The issue involves cerner millennium clinical notes® and affects users that utilize the product to create and/or edit textual notes.In cerner millennium®, when the user opens a patient's chart in the clinical notes® tab and then open another patient's chart, the previous patient's documents may be displayed on the note list.However, the correct patient information is displayed in the demographics bar.Patient care could be adversely affected, as clinicians could review documents for a patient other than the one they intended if the documents from the previous patient's chart are displayed.This issue could impact the treatment and care of the patient.Cerner has not received communication on any adverse patient events as a result of this issue.
 
Manufacturer Narrative
Cerner distributed a flash notification on (b)(6) 2020 to all potentially impacted client sites.The software notification includes a description of the issue and a software modification is being developed to address the issue for all sites that could be potentially impacted.Cerner distributed a flash notification on (b)(6) 2021 to all potentially impacted client sites.The software notification includes a description of the issue and a software modification has been developed to address the issue for all sites that could be potentially impacted.Cerner corporation considers this issue to be resolved and no further narrative is required for follow-up.
 
Event Description
The software product mentioned in this medwatch report may not be, by definition, a medical device; however, cerner has chosen to file this medwatch report in order to voluntarily notify the fda of a malfunction associated with this software product.The company's filing of this medwatch report does not signify cerner's belief or understanding that medical device reports are required to be filed for products such as cerner's millennium powerchart®, firstnet®, surgical powerchart® and discern® output viewer, nor are these products currently actively regulated by the fda.This report documents information related to an issue identified with functionality included in cerner's millennium powerchart®, firstnet®, surgical powerchart® and discern® output viewer.This issue occurs when users have patient's chart open and then open a different patient's chart.The new patient's information is correctly displayed on the demographics bar; however, the system may continue to display the previous patient's information.Patient care could be adversely affected if a clinician reviews patient data for the incorrect patient and makes clinical decisions for a different patient based on that information.Cerner has not received communication on any adverse patient events as a result of this issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CERNER'S MILLENNIUM POWERCHART®, FIRSTNET®, SURGICAL POWERCHART® AND DISCERN®
Type of Device
SOFTWARE
Manufacturer (Section D)
CERNER CORPORATION
8779 hillcrest road
kansas city MO 64138
Manufacturer (Section G)
CERNER CORPORATION
cerner corporation
8779 hillcrest road
kansas city MO 64138
Manufacturer Contact
shelley looby
8779 hillcrest road
kansas city, MO 64138
8162011368
MDR Report Key9765543
MDR Text Key193336609
Report Number1931259-2020-00004
Device Sequence Number1
Product Code LNX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Administrator/Supervisor
Remedial Action Modification/Adjustment
Type of Report Initial,Followup
Report Date 12/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2007.19 THROUGH 2018.09
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 07/05/2019
Initial Date FDA Received02/27/2020
Supplement Dates Manufacturer Received01/18/2019
Supplement Dates FDA Received12/17/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/03/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-