MAUDE Adverse Event Report: MEDTRONIC HEART VALVES DIVISION MOSAIC MITRAL BIOPROSTHETIC HEART VALVE; HEART-VALVE, NON-ALLOGRAFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC HEART VALVES DIVISION MOSAIC MITRAL BIOPROSTHETIC HEART VALVE; HEART-VALVE, NON-ALLOGRAFT TISSUE Back to Search Results
Model Number 310C29
Device Problems Gradient Increase (1270); Insufficient Information (3190)
Patient Problems Host-Tissue Reaction (1297); Inflammation (1932); Mitral Regurgitation (1964); Mitral Valve Stenosis (1965); Thrombus (2101)
Event Date 07/08/2019
Event Type  Injury  
Manufacturer Narrative
Product analysis: no product was returned.Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information that 7 months post implant of this 29mm bioprosthetic mitral valve, it was explanted and replaced with a 31mm non-medtronic bioprosthetic valve.The reasons for replacement were reported as severe stenosis with mild transvalvular regurgitation, pannus formation, and inflammation.It was also reported that the bioprosthetic valve appeared endothelialized with "organized thrombi on the undersurface of the leaflets".No additional adverse patient effects were reported.
 
Manufacturer Narrative
The device history record was reviewed and showed that this product met all manufacturing specifications for product released for distribution.No issues were identified that would have impacted this event.Review of the operative report and images submitted by the health care provider revealed no new information.Without the return of the valve for analysis, a root cause of the regurgitation, high gradients, stenosis, and clot could not be determined.B7: updated h6: coding updated.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOSAIC MITRAL BIOPROSTHETIC HEART VALVE
Type of Device
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
MDR Report Key9770002
MDR Text Key182004424
Report Number2025587-2020-00627
Device Sequence Number1
Product Code LWR
UDI-Device Identifier00643169594852
UDI-Public00643169594852
Combination Product (y/n)N
PMA/PMN Number
P990064
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup
Report Date 03/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date08/21/2023
Device Model Number310C29
Device Catalogue Number310C29
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/31/2020
Initial Date FDA Received02/28/2020
Supplement Dates Manufacturer Received02/25/2020
Supplement Dates FDA Received03/25/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age67 YR
-
-