• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MEDICAL CORPORATION AUTOCLAVABLE INTERNAL HANDLES (MSERIES); INTERNAL PADDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MEDICAL CORPORATION AUTOCLAVABLE INTERNAL HANDLES (MSERIES); INTERNAL PADDLE Back to Search Results
Model Number 8011-0501-01
Device Problem Failure to Deliver Shock/Stimulation (1133)
Patient Problem No Information (3190)
Event Type  malfunction  
Manufacturer Narrative
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
 
Event Description
Complainant alleged that while attempting to defibrillate a patient (age & gender unknown), the associated defibrillator failed to discharge using these internal handles.Complainant did not indicate that there was any adverse effect to the patient due to the reported malfunction.
 
Manufacturer Narrative
The customer was contacted for return of the suspect product.The product has not been returned to zoll for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOCLAVABLE INTERNAL HANDLES (MSERIES)
Type of Device
INTERNAL PADDLE
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key9770041
MDR Text Key181395727
Report Number1220908-2020-00479
Device Sequence Number1
Product Code LDD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K963781
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number8011-0501-01
Device Catalogue Number8011050101
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/07/2020
Initial Date FDA Received02/28/2020
Supplement Dates Manufacturer Received02/07/2020
Supplement Dates FDA Received03/30/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/01/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-