• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION STERILE FX25RWC W/ RES; BLOOD GAS OXYGENATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION STERILE FX25RWC W/ RES; BLOOD GAS OXYGENATOR Back to Search Results
Model Number 3CX*FX25RWC
Device Problem Obstruction of Flow (2423)
Patient Problem No Patient Involvement (2645)
Event Date 02/12/2020
Event Type  malfunction  
Manufacturer Narrative
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.(b)(4).
 
Event Description
The user facility reported to terumo cardiovascular that prior to cardiopulmonary bypass, during prime, the manifold line of the venous inlet was obstructed.No patient involvement.The product was changed out.Procedure was completed successfully.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on march 2, 2020.Upon further investigation of the reported event, the following information is new and/or changed: d4 (additional device information - added exp date).G4 (date received by manufacturer).G7 (indication that this is a follow-up report).H2 (follow-up due to additional information).H4 (device manufacture date).H6 (identification of evaluation codes 11, 3331, 4114, 3221, 4315).Method code #1: 11 - testing of device from same lot/batch retained by manufacturer.Method code #2: 3331 - analysis of production records.Method code #3: 4114 - device not returned.Results code: 3221 - no findings available.Conclusions code: 4315 - cause not established.The affected sample was not returned; therefore, a thorough investigation could not be conducted.A representative retention sample was reviewed and water was able to be passed through the manifold assembly without issue.All capiox manifolds are 100% leak tested, ensuring flow through the assembly, during production.All available information has been placed on file in quality management for appropriate tracking, trending and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERILE FX25RWC W/ RES
Type of Device
BLOOD GAS OXYGENATOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
MDR Report Key9777078
MDR Text Key200825259
Report Number1124841-2020-00053
Device Sequence Number1
Product Code DTZ
UDI-Device Identifier00699753450837
UDI-Public(01)00699753450837
Combination Product (y/n)N
PMA/PMN Number
K151791
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 04/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2021
Device Model Number3CX*FX25RWC
Device Catalogue NumberN/A
Device Lot NumberWM22
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/12/2020
Initial Date FDA Received03/02/2020
Supplement Dates Manufacturer Received04/07/2020
Supplement Dates FDA Received04/09/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-