Model Number N/A |
Device Problems
Difficult to Insert (1316); Material Deformation (2976)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Report source: (b)(6).Complaint sample was evaluated and the reported event was unable to confirmed.The device was not returned, however a thread like material which appears to be the barb that is used to lock the liner into the ring was returned.Dhr was reviewed and no discrepancies were found.Root cause was unable to be determined.A summary of the investigation has been sent to the complainant.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
It was reported that during a hip procedure, a liner was driven into a shell and was unable to be secured.Once the liner was taken out, a thread-like material came out allowing the liner to seat within the shell.There was no reported harm or injury to the patient.Attempts have been made and additional information on the reported event is unavailable at this time.
|
|