• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDIVANCE, INC. ¿ 1725056 ARCTICSUN GEL PADS; ARCTIC GEL PAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDIVANCE, INC. ¿ 1725056 ARCTICSUN GEL PADS; ARCTIC GEL PAD Back to Search Results
Catalog Number 318-02
Device Problems Patient-Device Incompatibility (2682); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Necrosis (1971); Injury (2348); Burn, Thermal (2530)
Event Type  Injury  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.
 
Event Description
It was reported that an infant patient developed fat necrosis to the skin during arctic sun therapy.Additional information was received from account manager on 18feb2020 that the pad had been discarded.The infant was noted to be (b)(6) at the time of treatment.The wound ostomy nurse had assessed the patient and a kenalog cream was applied to the injury.The complainant noted that the infant had been repositioned every four hours, per protocol.
 
Event Description
It was reported that an infant patient developed fat necrosis to the skin during arctic sun therapy.Additional information was received from account manager on (b)(6)2020 that the pad had been discarded.The infant was noted to be (b)(4) weeks at the time of treatment.The wound ostomy nurse had assessed the patient and a kenalog cream was applied to the injury.The complainant noted that the infant had been repositioned every four hours, per protocol.
 
Manufacturer Narrative
The reported event was inconclusive due to poor sample condition.Visual evaluation of the photo sample noted a physical reaction beneath/on the skin of a neonatal patient.The pad allegedly used to treat the patient was not in place or clearly visible on the patient.Without the physical sample or verification of the use of the pad on the affected area, it cannot be determined whether the device was related to the event or not.A potential root cause for the reported failure could be, ¿8.2.2 materials that are contacting the patient¿s intact skin are not biocompatible.¿ the lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "directions 1.Place the patient (1.8 - 4.5 kg; 4.0 - 9.9 lb) on the pad.Avoid placing the patients over the manifolds or other high pressure locations.The rate of temperature change and potentially the final achievable temperature is affected by pad surface area coverage, placement, patient size, and water temperature range.2.The pad surface must be contacting the skin for optimal energy transfer efficiency.A) if desired, the center section of the pad can be wrapped around the patient¿s torso and secured in place using the velcro tabs provided.¿ if this option is in use, ensure that the edges of the pad are away from articulating areas of the body to avoid irritation.¿ place pads to allow for full respiratory excursion.(e.G.Ensure free movement of the chest and abdomen are guaranteed).¿ the pads may be removed and reapplied if necessary.¿ pads should be placed on healthy, clean skin only.3.Due to the small patient size (1.8 - 4.5 kg; 4.0 - 9.9 lb) and the potential for rapid patient temperature change, it is recommended to use the following settings to the arctic sun® temperature management system: ¿ water temperature high limit: =40°c (104°f) ¿ water temperature low limit: =10°c (50°f) ¿ control strategy: 2 4.Due to the small patient size (1.8 - 4.5 kg; 4.0 - 9.9 lb) it is recommended to use the patient temperature high and patient temperature low alert settings.5.Place a core patient temperature probe and connect to the arctic sun® temperature management system patient temperature 1 connector for continuous patient temperature feedback.A rectal or esophageal temperature probe is recommended.6.Verify patient core temperature with an independent temperature probe before and at regular intervals during use.7.Attach the pad¿s line connectors to the fluid delivery line manifolds.8.See arctic sun® temperature management system operators manual and help screens for detailed instructions on system use.9.Begin treating the patient.10.If the pad fails to prime or a significant continuous air leak is observed in the pad return line, check the connections, then if needed, replace the leaking pad.Once the pad is primed, assure the steady state flow rate displayed on the control panel is appropriate.The minimum flow rate should be 1.1 l/m.11.When finished, purge water from pad." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCTICSUN GEL PADS
Type of Device
ARCTIC GEL PAD
Manufacturer (Section D)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
louisville CO 80027
MDR Report Key9780691
MDR Text Key188140200
Report Number1018233-2020-01486
Device Sequence Number1
Product Code DWJ
Combination Product (y/n)N
PMA/PMN Number
K142702
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Type of Report Initial,Followup
Report Date 04/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/03/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number318-02
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/13/2020
Date Manufacturer Received03/30/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Weight3
-
-