MAUDE Adverse Event Report: EDWARDS LIFESCIENCES, PR SWAN-GANZ BIPOLAR PACING CATHETER FOR SVC INSERTION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES, PR SWAN-GANZ BIPOLAR PACING CATHETER FOR SVC INSERTION Back to Search Results
Model Number PE074F5
Device Problem Pacing Problem (1439)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 02/13/2020
Event Type  malfunction  
Manufacturer Narrative
One bipolar pacing catheter with attached monoject 1.3 cc limited volume syringe was returned for evaluation.Continuity testing confirmed a full open condition in the proximal and distal circuits.A cut down of the catheter body was performed just proximal of the proximal electrode to expose the pacing lead wires.The proximal lead wire was found to be broken at the 1.3 cm area from the catheter tip and insulation was not present at the broken section.The distal lead wire was found to be broken around the solder joint.It was found that both the proximal and distal circuits were continuous from the broken section to the connector pin.The balloon inflated clear and concentric with 1.3 cc air and remained inflated for 5 minutes without leakage.No visible damage or defect to the catheter body, balloon, windings, or returned syringe was found.Balloon inflation test was performed using returned syringe with 1.3 cc air by holding the balloon under water.Visual examination was performed under microscope at 20x magnification and with the unaided eyes.A device history record review was completed and documented that device met all specifications upon distribution.Customer report of pacing issue was confirmed during the analysis.An investigation has been initiated to consider any potential manufacturing factors that may have contributed to this complaint and implement any necessary corrective actions.Swan-ganz pacing thermodilution (td) catheters serve as diagnostic and therapeutic tools in the management of critically ill patients.There are multiple failure modes that may require the exchange of a pacing catheter.Since proper functioning of the pacing catheter depends on the electrical continuity of its electrodes and internal wires, care should be exercised when handling the catheter.Stretching, kinking, or forceful wiping of the catheter may result in damage.After stable pacing has been confirmed, the proximal end of the catheter should be secured to the insertion site to prevent undue movement that could result in tip dislodgment and loss of capture, or catheter migration.Care should be taken not to kink the catheter body when securing it.In this complaint, it could not be determined if procedural factors or device handling may have contributed to the reported event.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
 
Event Description
It was reported that the swan ganz catheter was unable to pace during use.The catheter was exchanged and the problem was solved.It is unknown if the patient had a history of cardiac conduction defect.Patient demographic information requested but unavailable.There were no patient complications reported.
 
Manufacturer Narrative
Corrected data: f10, h6.Reference capa-20-00141.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN-GANZ BIPOLAR PACING CATHETER FOR SVC INSERTION
Type of Device
PACING CATHETER
Manufacturer (Section D)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR
MDR Report Key9788700
MDR Text Key194099052
Report Number2015691-2020-10798
Device Sequence Number1
Product Code LDF
Combination Product (y/n)N
PMA/PMN Number
K813521
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 02/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/04/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/26/2021
Device Model NumberPE074F5
Device Lot Number62503128
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/27/2020
Date Manufacturer Received07/23/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-