The reported field event of backup vvi was verified in the lab.However, the root cause of the reset could not be conclusively determined since the device was reprogrammed in the field and the event was not reproducible.Testing was performed and the device was found to be normal; telemetry, pacing, sensing, impedance, high voltage (hv) output, high voltage (hv) shock and patient notifier were tested on the bench.No anomaly was detected.The diaphragmatic stimulation could not be confirmed in the lab.The cause is most likely consistent with the lead positioning.The battery performance alert (bpa) was verified in the lab.However, further investigation showed the battery performance alert (bpa) was a false positive.Prior to the event, device code was reloaded in the field, which cleared the device¿s usage data as part of the procedure.The bpa algorithm uses device usage history data for calculations and since that data got cleared after reloading code, a false bpa was triggered.This is normal and expected behavior when the data is cleared and the device had been implanted for >7 years.Additional information: d10.
|