MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® I.C. COMPLETE CARE® FOLEY TRAY; FOLEY CATHETER (LATEX)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® I.C. COMPLETE CARE® FOLEY TRAY; FOLEY CATHETER (LATEX) Back to Search Results
Catalog Number 900014A
Device Problems Fitting Problem (2183); Detachment of Device or Device Component (2907)
Patient Problem No Patient Involvement (2645)
Event Date 02/10/2020
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.
 
Event Description
It was reported that the syringe came off easily from the inflation valve during a pretest.The device was not used.
 
Event Description
It was reported that the syringe came off easily from the inflation valve during a pretest.The device was not used.
 
Manufacturer Narrative
The reported event was unconfirmed since the problem could not be reproduced.The sample was evaluated and no abnormalities were observed that could lead to the reported event.Besides, no damage was observed on the returned valve.The device was tested using a lab syringe and once connected to the valve of returned catheter no difficulties in regards to connecting the syringe were observed to the valve of catheter.The catheter was able to be inflated and deflated without difficulties.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: ¿(1) do not inflate the balloon in the urethra.[the urethra may be injured.] (2) do not pull the catheter hard.[the bladder/urethra may be injured.] 2.Applicable patients (1) patients with delirium who might pull out catheter [when patient tugs at catheter unconsciously, the bladder and urethra may be damaged.] [contraindications] 1.Method for use: (1) do not reuse.(2) do not resterilize.(3) this device contains 10% povidone-iodine.For patients with past history of allergic hypersensitivity to povidone-iodine or iodine, consider using alternative disinfectants.(4) be careful that the catheter is not exposed to ointments, contrast medium or oil-based lubricants (including vegetable oils such as olive oil, mineral oils such as white petrolatum and animal oils).[they may damage the device and may burst balloon.] (5) do not hold the device with forceps, etc.Avoid contact with any blades or sharp-edged instruments.[catheter damage may cause balloon rupture and accidental balloon removal or failure to deflate or remove the balloon.] 2.Applicable patients (1) do not use in patients who are or have been allergic to natural rubber latex." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® I.C. COMPLETE CARE® FOLEY TRAY
Type of Device
FOLEY CATHETER (LATEX)
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key9794357
MDR Text Key185063837
Report Number1018233-2020-01573
Device Sequence Number1
Product Code MJC
Combination Product (y/n)N
PMA/PMN Number
K910318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,other
Type of Report Initial,Followup
Report Date 03/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/05/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/28/2022
Device Catalogue Number900014A
Device Lot NumberMYDU2295
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/21/2020
Date Manufacturer Received03/16/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-