MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION INFUSOR; PUMP, INFUSION, ELASTOMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION INFUSOR; PUMP, INFUSION, ELASTOMERIC Back to Search Results
Catalog Number J2C1711
Device Problem Improper Flow or Infusion (2954)
Patient Problem No Patient Involvement (2645)
Event Date 02/06/2020
Event Type  malfunction  
Manufacturer Narrative
The device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that there was an air bubble observed within the tubing of a small volume infusor.The air bubble was discovered at the hospital prior to patient use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Additional information was added: the device was manufactured from august 5, 2019 - august 6, 2019.The actual sample was received for evaluation.Visual inspection was performed using the naked eye which revealed a clear particle, measured 0.20 square mm.The particle was not in the fluid path because it was embedded in the material of the tubing line.The particle was subsequently identified by fourier-transform infrared spectroscopy to be polyvinyl chloride (pvc).Pvc is a raw material of the tubing line.The reported condition of particulate matter was verified on the returned sample.The cause of the condition is related to a supplier manufacturing issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
building 3
irvine CA 92614
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key9801939
MDR Text Key182428653
Report Number1416980-2020-01259
Device Sequence Number1
Product Code MEB
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date 04/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/06/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberJ2C1711
Device Lot Number19H004
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/18/2020
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/30/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-