• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, BRASIL LTDA. (PAMPULHA) EPIC STENTED PORCINE HEART VALVE W/FLEXFIT SYSTEM; HEART-VALVE, NON-ALLOGRAFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, BRASIL LTDA. (PAMPULHA) EPIC STENTED PORCINE HEART VALVE W/FLEXFIT SYSTEM; HEART-VALVE, NON-ALLOGRAFT TISSUE Back to Search Results
Model Number E100-29M
Device Problems Backflow (1064); Material Integrity Problem (2978); Material Split, Cut or Torn (4008)
Patient Problems Mitral Insufficiency (1963); Hemolytic Anemia (2279)
Event Date 02/20/2020
Event Type  Injury  
Manufacturer Narrative
An event of insufficiency due to a torn cusp was reported.The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications at the time of commercialization.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
On (b)(6) 2017, a 29mm epic valve was selected for implant.On (b)(6) 2020, the valve was explanted after echocardiogram revealed the patient had severe mitral insufficiency after she became admitted due to hemolytic anemia.Upon explant, one of the leaflets had a tear.A replacement valve was implanted and the patient was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIC STENTED PORCINE HEART VALVE W/FLEXFIT SYSTEM
Type of Device
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
ST. JUDE MEDICAL, BRASIL LTDA. (PAMPULHA)
rua professor jose vieira de mendonça 1301
engenho nogueira - belo horizonte - mg
belo horizonte 31310 -260
BR  31310-260
Manufacturer (Section G)
ST. JUDE MEDICAL, BRASIL LTDA. (PAMPULHA)
rua professor jose vieira de mendonça 1301
engenho nogueira - belo horizonte - mg
belo horizonte 31310 -260
BR   31310-260
Manufacturer Contact
pamela yip
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key9803853
MDR Text Key187624435
Report Number3001883144-2020-00017
Device Sequence Number1
Product Code LWR
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
P040021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 03/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/08/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/25/2019
Device Model NumberE100-29M
Device Catalogue NumberE100-29M
Device Lot NumberBR00004581
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/22/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/26/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Life Threatening; Other; Required Intervention;
-
-