• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DERUNGS LICHT AG VISIANO 10-1; DEVICE, MEDICAL EXAMINATION, AC POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DERUNGS LICHT AG VISIANO 10-1; DEVICE, MEDICAL EXAMINATION, AC POWERED Back to Search Results
Model Number P S10
Device Problems Melted (1385); Material Deformation (2976)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/09/2019
Event Type  malfunction  
Event Description
Clinical engineering technician notes - "i found 10 out of 22 giraffe incubator/warmer exam lights, which are mounted on most giraffe that were melted and deformed due to being left under the radiant warmer.In some cases the exam light was melted so bad that the inside of the lamp was exposed.Another concern is that the fumes from melting plastic can be harmful to patients.I found 2 of 7 in one nicu room were melted, 5 of 10 in another room were melted, 2 of 4 in one intensive care unit were melted, and i found one in the 2nd floor storage area (by the pharmacy) that was melted.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISIANO 10-1
Type of Device
DEVICE, MEDICAL EXAMINATION, AC POWERED
Manufacturer (Section D)
DERUNGS LICHT AG
9 w century dr
wheeling IL 60090
MDR Report Key9805099
MDR Text Key182556011
Report Number9805099
Device Sequence Number1
Product Code KZF
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/10/2020,01/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberP S10
Device Catalogue NumberD15781000
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/10/2020
Event Location Hospital
Date Report to Manufacturer03/09/2020
Initial Date Manufacturer Received Not provided
Initial Date FDA Received03/09/2020
Type of Device Usage N
Patient Sequence Number1
-
-