Model Number SN60WF
Device Problem Material Deformation (2976)
Patient Problem No Code Available (3191)
Event Date 02/04/2020
Event Type  Injury  
Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.Additional information has been requested.The manufacturer internal reference number is: (b)(4).
 
Event Description
A surgical tech reported that during an intraocular lens (iol) implant procedure, the lens was inserted into the patient's eye and it was observed that the haptic was slightly kinked to where it wasn't sitting properly.The iol was cut out and a new lens was implanted.An enlarged incision was required during the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ NATURAL SINGLEPIECE IOL
Type of Device
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8175514979
MDR Report Key9805863
MDR Text Key182561075
Report Number1119421-2020-00404
Device Sequence Number1
Product Code HQL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P930014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSN60WF
Device Catalogue NumberSN60WF.195
Device Lot Number12739510
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/13/2020
Initial Date FDA Received03/09/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/27/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MONARCH III D CARTRIDGES; MONARCH III IOL DELIVERY SYST; UNSPECIFIED PROVISC
Patient Outcome(s) Required Intervention;
-
-