Catalog Number 4251644-01 |
Device Problem
Protective Measures Problem (3015)
|
Patient Problem
Needle Stick/Puncture (2462)
|
Event Date 01/14/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
This report has been identified as b.Braun melsungen (b)(4) internal report # (b)(4).Within the current restriction for covid-19 the sample cannot be sent for investigation.Without the actual sample (or device), a thorough investigation could not be performed and no specific conclusion can be drawn as to the cause of the reported event.Device history record review (dhr): reviewed the device history record and no abnormalities found during in process and final control inspection.If additional pertinent information becomes available, a follow up report will be submitted.Note: this report is being filed for an item number that is not sold in the united states, however similar items are sold in the united states by b.Braun medical, inc.
|
|
Event Description
|
As reported by the user facility (translation of user facility information by bbm sales organization in (b)(4)): faulty product - biological risk.After having placed a venous access in the patient, after having extracted the mandrin from the cannula, the operator punctured himself with the tip of the mandrin which was not covered with the automatic cover system of the tip that even if it was automatically snapped, it did not covered the needle tip.Consequence: operator puncture, with risk of biological contamination.
|
|