Catalog Number 4251644-01
Device Problem Protective Measures Problem (3015)
Patient Problem Needle Stick/Puncture (2462)
Event Date 01/14/2020
Event Type  malfunction  
Manufacturer Narrative
This report has been identified as b.Braun melsungen (b)(4) internal report # (b)(4).Within the current restriction for covid-19 the sample cannot be sent for investigation.Without the actual sample (or device), a thorough investigation could not be performed and no specific conclusion can be drawn as to the cause of the reported event.Device history record review (dhr): reviewed the device history record and no abnormalities found during in process and final control inspection.If additional pertinent information becomes available, a follow up report will be submitted.Note: this report is being filed for an item number that is not sold in the united states, however similar items are sold in the united states by b.Braun medical, inc.
 
Event Description
As reported by the user facility (translation of user facility information by bbm sales organization in (b)(4)): faulty product - biological risk.After having placed a venous access in the patient, after having extracted the mandrin from the cannula, the operator punctured himself with the tip of the mandrin which was not covered with the automatic cover system of the tip that even if it was automatically snapped, it did not covered the needle tip.Consequence: operator puncture, with risk of biological contamination.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTROCAN SAFETY®
Type of Device
CATHETER,INTRAVASCULAR,THE
Manufacturer (Section D)
B. BRAUN MELSUNGEN AG
carl-braun-str. 1
melsungen,
Manufacturer (Section G)
B. BRAUN MELSUNGEN AG
carl-braun-str. 1
melsungen,
Manufacturer Contact
jonathan severino
861 marcon blvd.
allentown, PA 18109
4842408332
MDR Report Key9805904
MDR Text Key196307370
Report Number9610825-2020-00050
Device Sequence Number1
Product Code FOZ
Combination Product (y/n)N
Reporter Country CodeIT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/09/2020
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/01/2023
Device Catalogue Number4251644-01
Device Lot Number18N03G8301
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/05/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/03/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-