• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND ATTAIN COMMAND; CATHETER, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND ATTAIN COMMAND; CATHETER, PERCUTANEOUS Back to Search Results
Model Number 6250VI-3D
Device Problem Material Puncture/Hole (1504)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/11/2019
Event Type  malfunction  
Manufacturer Narrative
Delivery catheter was returned and analyzed.Analysis indicated the mechanical wire had a penetrated lumen wall.The mechanical operation of the catheter shaft was kinked and buckled.The mechanical operation of the catheter shaft was bent.Visual analysis of the lead indicated damage during use.The analyst noted that the catheter was returned with the tip was damaged.Dried blood is visible on the catheter.Visual ispection showed damaged during use.The dilator was returned with the catheter shaft.If information is provided in the future, a supplemental report will be issued.
 
Event Description
The catheter was replaced during the implant procedure following difficulty inserting into the sheath and the tip of the catheter being crushed.The catheter was returned to the manufacturer, analyzed, and tested out of specification.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTAIN COMMAND
Type of Device
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
IE 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
IE  
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key9806058
MDR Text Key191621802
Report Number9612164-2020-01084
Device Sequence Number1
Product Code DQY
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K123153
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 03/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/03/2021
Device Model Number6250VI-3D
Device Catalogue Number6250VI-3D
Device Lot Number0009953645
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/27/2020
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/14/2020
Initial Date FDA Received03/09/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/04/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-