ROCHE DIAGNOSTICS ELECSYS FT4 III ASSAY; RADIOIMMUNOASSAY, FREE THYROXINE Back to Search Results
Model Number FT4 G3
Device Problem High Test Results (2457)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/20/2020
Event Type  malfunction  
Manufacturer Narrative
The investigation is ongoing.This event occurred in (b)(6).(b)(4).
 
Event Description
The initial reporter received questionable elecsys ft4 iii and elecsys ft3 iii assay results for one patient from cobas e 801 module serial number (b)(4).Sample from the patent was submitted for initial investigation and was tested on centaur, cobas e801, cobas e602, and cobas e411 analyzers.Refer to the attachment to the medwatch for all patient data.Refer to the medwatch with patient identifier (b)(4) for the ft3 iii assay.The customer reported out the results to a physician.
 
Manufacturer Narrative
Sample from the patient was submitted for investigation.An interfering factor to streptavidin was identified in the sample.This interference was documented in product labeling for the assay: in rare cases, interference due to extremely high titers of antibodies to analyte-specific antibodies, streptavidin or ruthenium can occur.These effects are minimized by suitable test design.No product problem could be found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS FT4 III ASSAY
Type of Device
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
MDR Report Key9810586
MDR Text Key220573557
Report Number1823260-2020-00689
Device Sequence Number1
Product Code CEC
Combination Product (y/n)N
PMA/PMN Number
K131244
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 04/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/10/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFT4 G3
Device Catalogue Number07976887190
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Date Manufacturer Received02/20/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-