ADVANCED STERILIZATION PRODUCTS STERRAD 100NX STERILIZER 1-DR W/ DUO; STERRAD EQUIPMENT Back to Search Results
Model Number 10104-003
Device Problem Smoking (1585)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/14/2020
Event Type  malfunction  
Manufacturer Narrative
A field service engineer was dispatched to the customer site and identified the oil mist filter retaining screw was loose.The screw was tightened to resolve the smoke/haze issue.Unit meets specifications and was returned to service.Asp complaint ref #: (b)(4).
 
Event Description
A customer reported an event of a smoke or haze emitting from the sterrad® 100nx sterilizer.There was no report of any injuries or human reactions.The customer was advised to power down unit and avoid the area.An asp field service engineer was dispatched to assess the unit onsite.This event is being reported as a malfunction report subsequent to a serious injury.
 
Manufacturer Narrative
H3: asp investigation summary: the investigation included a review of the device history record (dhr), trending analysis of the smoke/haze issue and system risk analysis (sra).The dhr was reviewed and no issues relating the failure mode were noted.The involved unit met manufacturer specifications at the time of release.Trending analysis of the smoke/haze issue was reviewed from to and no significant trend was observed.The sra shows the risk for exposure to toxic or corrosive material to be "low." no parts were replaced to resolve the issue.The assignable cause of the smoke/haze issue is likely due to the oil mist filter.The field service engineer tightened/adjusted the part and confirmed the sterrad® 100nx was restored to proper function after service.The issue was resolved at the customer facility.Asp will continue to track and trend this issue.Asp complaint ref #: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERRAD 100NX STERILIZER 1-DR W/ DUO
Type of Device
STERRAD EQUIPMENT
Manufacturer (Section D)
ADVANCED STERILIZATION PRODUCTS
33 technology drive
irvine CA
MDR Report Key9812456
MDR Text Key185486170
Report Number2084725-2020-50022
Device Sequence Number1
Product Code MLR
UDI-Device Identifier10705037047581
UDI-Public10705037047581
Combination Product (y/n)N
PMA/PMN Number
K071385
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 08/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10104-003
Device Catalogue Number10104003
Was the Report Sent to FDA? Yes
Initial Date Manufacturer Received 02/14/2020
Initial Date FDA Received03/10/2020
Supplement Dates Manufacturer Received08/11/2020
Supplement Dates FDA Received08/12/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-