• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. REMSTAR; VENTILATOR, NON-CONTINUOUS (RESPIRATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. REMSTAR; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number 150P
Device Problems Mechanical Problem (1384); Decrease in Pressure (1490)
Patient Problem High Blood Pressure/ Hypertension (1908)
Event Date 01/20/2020
Event Type  Injury  
Event Description
A complainant contacted the manufacturer alleging the continuous positive airway pressure (cpap) device he was using caused a hypertensive episode due to inadequate pressure being delivered.No immediate medical intervention was required.The event reportedly lasted two days.He was prescribed medication for hypertension at a later date, and is reportedly using an old device he still has in his possession.He stated he has not had any further issues with hypertension.The manufacturer requested return of the device for investigation.The complainant refused to return it to the manufacturer.The durable medical equipment (dme) provider listed in the equipment record as the distributor of the device was contacted and informed the manufacturer the device was provided to a patient with a different last name, and they have no record of the complainant having this device.The device was sold to the customer in 2010.The complainant stated he was "given" the device 6 or 7 years ago and had taken it to his physician's office to be checked and have the pressure set to 11 cm h2o.The complainant also stated that he took the device to a different dme recently to have the pressure checked, and that two manometers measured the pressure to be within specification of the prescribed pressure.The complainant stated he still didn't think the device was providing the correct pressure even though the manometers showed it was.A search in the manufacturer's database indicates this device has not been returned for any service since date of manufacturer in 2010.This device is no longer serviceable.Patient labeling warns the user that if they detect any unexplained changes in the performance of the unit, if the unit is dropped or mishandled, if liquid or water is spilled into the enclosure or if the enclosure is broken, to stop using the device, unplug it and seek assistance from respironics or an authorized service center.The intended use of this device is only for the treatment of obstructive sleep apnea in spontaneously breathing patients weighing more than 66 pounds.The short-term loss of therapy for this patient class does not pose a significant health or safety risk.As the complainant refuses to return the device for investigation, the manufacturer is unable to confirm the allegation that inadequate cpap pressure caused or contributed to the complainant's hypertensive episode.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REMSTAR
Type of Device
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville, pa
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville, pa
Manufacturer Contact
adam price
1001 murry ridge lane
murrysville, pa 
3349303
MDR Report Key9818394
MDR Text Key182949949
Report Number2518422-2020-00496
Device Sequence Number1
Product Code BZD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K091319
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
Report Date 02/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number150P
Device Catalogue Number150P
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/18/2020
Initial Date FDA Received03/11/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/23/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age73 YR
-
-