MEDTRONIC EUROPE SARL VIVA¿ XT CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO Back to Search Results
Model Number DTBA2D4
Device Problem Failure to Capture (1081)
Patient Problem Cardiac Arrest (1762)
Event Date 02/11/2020
Event Type  Injury  
Manufacturer Narrative
Concomitant medical products: product id: 6935m62, serial#: (b)(4).Product id: 383069, serial#: (b)(4), implanted: (b)(6) 2019.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that approximately one month following implant, the patient was experiencing symptoms and a holter test was performed.It was noted the right ventricular (rv) lead output had been programmed subthreshold with left ventricular (lv) lead only pacing through his bundle pacing.Asystole was observed on the holter and it was added the device was utilizing rv only pacing during atrial capture management testing, however, the rv subthreshold programming resulted in no ventricular pacing support during the time of testing.The rv output was programmed off and the lv output was reprogrammed to a lower value, which improved the capture threshold in clinic.The rv lead and lv lead remain in use.No further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIVA¿ XT CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
CH  1131
Manufacturer (Section G)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
CH   1131
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key9821993
MDR Text Key183517196
Report Number9614453-2020-00905
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P010031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/11/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/28/2019
Device Model NumberDTBA2D4
Device Catalogue NumberDTBA2D4
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/11/2020
Date Device Manufactured06/01/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Life Threatening; Required Intervention;
-
-