Model Number 816570 |
Device Problem
Protective Measures Problem (3015)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The field service representative (fsr) confirmed that the roller pump did not respond to the low level alarm due to the fact that the level sensor was not plugged into the alarm port on the level pod.There was an alert sensor plugged into the alert port on the level pod.The perfusionist wanted the alert sensor plugged into the alarm port in the level pod.The pump responded correctly to the low level alarm.The unit operated to the manufacturer's specifications.
|
|
Event Description
|
It was reported that the roller pump was not responding to alarms.No other details regarding the nature of this event were provided.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|