• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS- ADVANCED PERFUSION SYSTEM 1

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS- ADVANCED PERFUSION SYSTEM 1 Back to Search Results
Model Number 816570
Device Problem Protective Measures Problem (3015)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The field service representative (fsr) confirmed that the roller pump did not respond to the low level alarm due to the fact that the level sensor was not plugged into the alarm port on the level pod.There was an alert sensor plugged into the alert port on the level pod.The perfusionist wanted the alert sensor plugged into the alarm port in the level pod.The pump responded correctly to the low level alarm.The unit operated to the manufacturer's specifications.
 
Event Description
It was reported that the roller pump was not responding to alarms.No other details regarding the nature of this event were provided.
 
Manufacturer Narrative
The reported complaint was confirmed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS- ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key9825580
MDR Text Key183385159
Report Number1828100-2020-00120
Device Sequence Number1
Product Code DTQ
UDI-Device Identifier00886799001363
UDI-Public(01)00886799001363(11)180417
Combination Product (y/n)N
PMA/PMN Number
K172220
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 05/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/12/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number816570
Device Catalogue Number816570
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/01/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-