Advanced bionics considers the investigation into this reportable event as closed.The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.The external visual inspection revealed silicone slices on the top cover prior to receipt.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The device passed the electrical and mechanical tests performed.Advanced electrical testing performed on the device showed lower than typical range on some of the electrodes.It is the opinion of advanced bionics explant review board that some of these electrode test results should be considered as failures.The failure of this device is attributed to an electrode short in the electrode pocket.A corrective action was initiated.This version of the ultra device is no longer distributed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|