MAUDE Adverse Event Report: EDWARDS LIFESCIENCES DR PRESSURE MONITORING KIT; TRANSDUCER, PRESSURE, CATHETER TIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES DR PRESSURE MONITORING KIT; TRANSDUCER, PRESSURE, CATHETER TIP Back to Search Results
Model Number PXVP2260
Device Problems Disconnection (1171); Detachment of Device or Device Component (2907)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 02/17/2020
Event Type  malfunction  
Manufacturer Narrative
The device evaluation is anticipated.However, the complaint cannot not be confirmed without the completion of a product evaluation.A supplemental report will be forthcoming with the evaluation and device history results when received.
 
Event Description
It was reported that with this pressure monitoring set with vamp, the pressure line got disconnected from the sampling site, while priming.There was no allegation of patient injury.The device was available for evaluation.Patient demographics are unavailable.
 
Manufacturer Narrative
It was reported that with a pressure motoring set with vamp, the pressure line got disconnected from the sampling site during priming.  further follow up with the customer indicated that the device would not be returned for evaluation; it was discarded at the hospital due to the covid-19 pandemic.Without return of the unit it is not possible to determine if some damage or defect existed on the unit that could have contributed to the event.It is not known if some procedural factors may have contributed to the event.No corrective actions will be taken at this time.A review of the manufacturing records indicated that the product met specifications upon release.  it is common clinical practice to inspect all products before usage.These products are used by highly trained clinicians, experienced in identifying and mitigating any hazards that arise during use.In addition, they are used in critical care units or ors where patients are closely monitored.If the pressure tubing becomes detached during use, it will affect the pressure waveform, which will immediately alert the clinician to begin the troubleshooting process.In this case, there were no patient complications noted.   complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
 
Manufacturer Narrative
Reference capa-20-00141.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRESSURE MONITORING KIT
Type of Device
TRANSDUCER, PRESSURE, CATHETER TIP
Manufacturer (Section D)
EDWARDS LIFESCIENCES DR
parque industrial de itabo
carr. sanchez km 18.5
haina, san cristobal
DR 
MDR Report Key9825899
MDR Text Key194508921
Report Number2015691-2020-10933
Device Sequence Number1
Product Code DXO
Combination Product (y/n)N
PMA/PMN Number
K925638
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 02/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/09/2021
Device Model NumberPXVP2260
Device Catalogue NumberPXVP2260
Device Lot Number62538060
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/25/2020
Initial Date FDA Received03/12/2020
Supplement Dates Manufacturer Received03/31/2020
07/23/2020
Supplement Dates FDA Received04/13/2020
01/09/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age84 YR
Patient Weight64
-
-