The field service engineer could not determine a cause.Rinse mechanism levels and pump pressures were checked, with no abnormalities noted.Precision studies were performed and all values recovered within guidelines.The customer ran calibration and controls.Controls recovered within range.No alarms were noted on the last calibration performed on (b)(6) 2020.Quality control data was evaluated and controls were within range, showing no indication of a reagent or instrument performance issue.The investigation could not identify a product problem.The cause of the event could not be determined.
|
The initial reporter stated they received discrepant results for two patient samples tested with iron2 iron gen.2 on a cobas 8000 c 502 module.On (b)(6) 2020, controls passed for the iron reagent pack that was in use.On (b)(6) 2020, controls failed on the reagent pack and it had to eventually be replaced.It was noted that the complained reagent pack was near the end of its test volume.The customer repeated patient samples tested on (b)(6) 2020 and two samples had discrepant results.The initial values from these samples were reported outside of the laboratory.The repeat values were believed to be correct and corrected reports were issued.The first sample initially resulted with an iron value of 33 ug/dl and repeated as 19 ug/dl.The second sample, from a male patient, initially resulted with an iron value of 30 ug/dl and this repeated as 13 ug/dl.The c 502 analyzer serial number is (b)(4).
|