• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GII UNIVERSAL EXTRACTOR; PROSTHESIS, KNEE, FMOROTIBIAL, SEMI-CONSTRAIND, CEMENTED, METAL/POLYMR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GII UNIVERSAL EXTRACTOR; PROSTHESIS, KNEE, FMOROTIBIAL, SEMI-CONSTRAIND, CEMENTED, METAL/POLYMR Back to Search Results
Model Number 71440366
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/27/2020
Event Type  malfunction  
Event Description
It was reported that during a tka procedure the genesis ii universal extractor broke at weld while extracting the femoral cutting block.Nothing was left in the patient.The case was finished as planned because a backup device was not needed.
 
Manufacturer Narrative
The associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The handles/blocks extractor end of the device fractured off, rendering the device inoperable.The piece that fractured off was not returned.The device was manufactured in 2014 and shows signs of extensive wear/ usage.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII UNIVERSAL EXTRACTOR
Type of Device
PROSTHESIS, KNEE, FMOROTIBIAL, SEMI-CONSTRAIND, CEMENTED, METAL/POLYMR
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key9829190
MDR Text Key183288992
Report Number1020279-2020-00860
Device Sequence Number1
Product Code HRY
UDI-Device Identifier03596010215192
UDI-Public03596010215192
Combination Product (y/n)N
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 07/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/13/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71440366
Device Catalogue Number71440366
Device Lot Number14JM15503
Was Device Available for Evaluation? Yes
Date Manufacturer Received07/06/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-