• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD ENDO CLIP III; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD ENDO CLIP III; CLIP, IMPLANTABLE Back to Search Results
Model Number 176630
Device Problem Mechanics Altered (2984)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, the surgeon was able to squeeze the handle, there were issues experienced with the loading or firing of the clips, the clips were not able to be fired.There was no patient involvement.
 
Manufacturer Narrative
H3 evaluation summary: post market vigilance (pmv) led an evaluation of one device.The instrument was received partially applied with eight clips remaining.No visual abnormalities were observed with the instrument.The instrument was applied to appropriate test media for functional evaluation.The instrument was found to cycle without binding.Clips advanced into the jaws, formed properly, and were held securely in place after full formation was achieved and the firing handle was released.In addition, when the cartridge was empty, the interlock engaged to prevent the jaws from approximating.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO CLIP III
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO  0101
MDR Report Key9834866
MDR Text Key183656519
Report Number9612501-2020-00462
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10884521057852
UDI-Public10884521057852
Combination Product (y/n)N
PMA/PMN Number
K100242
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 05/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/16/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2021
Device Model Number176630
Device Catalogue Number176630
Device Lot NumberJ8G0186X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/16/2020
Date Manufacturer Received05/05/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-