According to the reporter, when clipping the gallbladder during a laparoscopic cholecystectomy, the jaws would not open.The central side was sealed with an energy device and the device was removed from the tissue by additionally resecting the tissue.The procedure was completed with another device.
|
Post market vigilance (pmv) led an evaluation of one device.Microscopic inspection of the instrument noted presence of clips in the tube.However, it was observed that the firing handle was broken in a manner consistent with cycling through the end of application safety interlock.The red indicator was visible in the window.The tissue stop was disengaged.Functional evaluation could not be performed as the safety interlock feature was engaged.Disassembly of the instrument confirmed the firing handle had broken in a manner consistent with cycling through the end of application safety interlock, and the pusher bar was slightly buckled.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the buckled pusher bar and jammed clip condition may occur under the following conditions.If the load indicator printed on the distal end of the instrument shaft has not cleared the distal end of the cannula and the instrument is fired.If a heavy load is applied to the side of the jaw which partially closes it during the clip loading process.Replication of the broken handle (broken interlock) may occur when an attempt has been made to cycle the handle after it has engaged in safety interlock due to the buckled pusher bar.The safety interlock feature engages to prevent the jaws from closing in the absence of a clip.The firing handle will fail, as observed, if an attempt is made to over-ride the safety interlock.The root cause of the observed damage was due to the product not being medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|