Model Number 8888145016
Device Problem Material Integrity Problem (2978)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/25/2020
Event Type  malfunction  
Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led a photographic evaluation of one device.The visual inspection of the returned photo noted: the image depicts the top of the blue luer adapter looking down.The threads are mangled and the adapter has a crack.Without the physical device functional evaluation is precluded.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during use, on preparation for dialysis, a crack/split was noted on the blue luer adapter, thus affecting normal use.There was no leak, no cleaning agent was used, tego was not utilized, no instrument was used to tighten or loosen the device.There was no blood loss and no transfusion was required.No other products was utilized with the device.The catheter was repaired and the adapter was replaced to resolve the issue.Iodine was the cleaning agent used.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PALINDROME
Type of Device
CATHETER, HEMODIALYSIS, IMPLANTED
Manufacturer (Section D)
COVIDIEN MFG SOLUTIONS S.A.
edificio b20, calle #2
alajuela 20101
CR  20101
Manufacturer (Section G)
COVIDIEN MFG SOLUTIONS S.A.
edificio b20, calle #2
alajuela 20101
CR   20101
Manufacturer Contact
lisa hernandez
15 hampshire street
mansfield, MA 02048
2034925563
MDR Report Key9835789
MDR Text Key194764813
Report Number3009211636-2020-00069
Device Sequence Number1
Product Code MSD
UDI-Device Identifier20884521013176
UDI-Public20884521013176
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K111372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/16/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/20/2022
Device Model Number8888145016
Device Catalogue Number8888145016
Device Lot Number1729100103
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/28/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/20/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-