US SURGICAL PUERTO RICO ENDO CLINCH II; FORCEPS, OBSTETRICAL Back to Search Results
Model Number 174317
Device Problem Material Split, Cut or Torn (4008)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/24/2020
Event Type  malfunction  
Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led a photographic evaluation of one device.A visual inspection of the returned photo(s) noted that the plastic sheath along the shaft had shearing damage and was peeling.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the reported condition may occur where an excessive force or leverage is applied to the device while retracting the clevis assembly and the clevis boot is subsequently damaged.This subsequent damage may result in improper retraction and deployment of the jaws during articulation and may cause shearing of the plastic sheath.The root cause of the observed damage was found to be due to the device not being used as intended which caused or contributed to the reported condition.No further actions have been deemed necessary at this time.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during laparoscopic cholecystectomy, the surgeon put the instrument down the port and noticed the insulation torn on the distal end of the shaft of the instrument.Another instrument was pulled and used to complete the case.There was no patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO CLINCH II
Type of Device
FORCEPS, OBSTETRICAL
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer (Section G)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer Contact
lisa hernandez
60 middletown ave.
north haven, CT 06473
2034925563
MDR Report Key9835984
MDR Text Key183538830
Report Number2647580-2020-00966
Device Sequence Number1
Product Code HDA
UDI-Device Identifier10884523000832
UDI-Public10884523000832
Combination Product (y/n)N
Reporter Country CodeEI
PMA/PMN Number
K903205
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/16/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2024
Device Model Number174317
Device Catalogue Number174317
Device Lot NumberP9E1274Y
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/28/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/23/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-