Model Number EL5ML
Device Problems Failure to Form Staple (2579); Material Twisted/Bent (2981); Device Fell (4014)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/14/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # t94n61 investigation summary: the analysis results found that the el5ml device was returned with no apparent damage.In an attempt to replicate the reported incident, the instrument was tested for functionality.During the analysis, the device was cycled and it fed and formed five conforming clips.The event described could not be confirmed as the device performed without any difficulties noted.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
 
Manufacturer Narrative
(b)(4).Batch # unk.Attempts have been made to retrieve the device.To date, the device has not been returned.If the device or further details are received at a later date, a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformance were identified.
 
Event Description
It was reported that during an unknown procedure, the clip applier pre-loaded on the tip and was ready to fire, but after the surgeon pulled the trigger, the clip legs got twisted on the vessel and fell off.There were no patient consequences.No additional information is available at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGAMAX-5MM ENDO CLIP APPLIER
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
kara ditty-bovard
475 calle c
guaynabo 00969
6107428552
MDR Report Key9836281
MDR Text Key189760355
Report Number3005075853-2020-01607
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705036001843
UDI-Public10705036001843
Combination Product (y/n)N
PMA/PMN Number
K050344
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEL5ML
Device Catalogue NumberEL5ML
Device Lot NumberT94T8A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/27/2020
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/18/2020
Initial Date FDA Received03/16/2020
Supplement Dates Manufacturer Received03/27/2020
Supplement Dates FDA Received04/24/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/20/2019
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-