Model Number EL5ML |
Device Problems
Failure to Form Staple (2579); Material Twisted/Bent (2981); Device Fell (4014)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 02/14/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Batch # t94n61 investigation summary: the analysis results found that the el5ml device was returned with no apparent damage.In an attempt to replicate the reported incident, the instrument was tested for functionality.During the analysis, the device was cycled and it fed and formed five conforming clips.The event described could not be confirmed as the device performed without any difficulties noted.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
|
|
Manufacturer Narrative
|
(b)(4).Batch # unk.Attempts have been made to retrieve the device.To date, the device has not been returned.If the device or further details are received at a later date, a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformance were identified.
|
|
Event Description
|
It was reported that during an unknown procedure, the clip applier pre-loaded on the tip and was ready to fire, but after the surgeon pulled the trigger, the clip legs got twisted on the vessel and fell off.There were no patient consequences.No additional information is available at this time.
|
|