Catalog Number 122432254
Device Problem Device Dislodged or Dislocated (2923)
Patient Problems Joint Dislocation (2374); No Code Available (3191)
Event Type  Injury  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
The primary surgery was performed via tha (date of surgery was unknown).It was reported that since reduction was performed under fluoroscopy to the patient who was under general anesthesia however it was not possible, the revision surgery was performed on (b)(6) 2020 by replacing the liner (p/n: 122432254), the head (p/n: 136532320) due to dislocation.The surgery was completed, and it was unknown whether there was a surgical delay or not.The patient mentioned that probably her hip joint was twisted when she was crouched down.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint (b)(4).Investigation summary : no device associated with this report was received for examination.A worldwide lot specific complaint database search, or device history record (dhr) review was not possible because the required lot code(s) was not provided.Based on previous investigations this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Corrective action was not indicated.The complaint was received into the company with the following comment: the primary surgery was performed via tha (date of surgery was unknown).It was reported that since reduction was performed under fluoroscopy to the patient who was under general anesthesia however it was not possible, the revision surgery was performed on (b)(6) 2020 by replacing the liner (p/n: 122432254), the head (p/n: 136532320) due to dislocation.The surgery was completed, and it was unknown whether there was a surgical delay or not.The patient mentioned that probably her hip joint was twisted when she was crouched down for offering flowers to buddhist altar on her husband death day, so dislocation occurred.No further information is available.No device associated with this report was received for examination.A worldwide lot specific complaint database search, or device history record (dhr) review, was not possible because the required lot code(s) was not provided.The reported event is considered one of the possible complications of joint replacement.The complaint information provided has been reviewed for complaint coding, medical device reporting, and other data required by the complaint system.Follow-up for additional event information, if applicable, was conducted utilizing work instruction wi-7915 appendix a.Without the physical complaint sample(s) associated with this report, it was not possible to determine if the device(s) failed to meet specification(s) at the time it was released for distribution.The device(s) associated with this event were used in the treatment of the patient as prescribed by the presiding surgeon.From the event information received, it was not possible to determine the relationship of the device to the reported event.No information received with this individual complaint indicates that a broader investigation or corrective action was necessary.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN MAR ETO LIP LNR 32IDX54OD
Type of Device
PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key9841388
MDR Text Key183717449
Report Number1818910-2020-08156
Device Sequence Number1
Product Code JDJ
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 02/19/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/17/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number122432254
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Date Manufacturer Received03/27/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-