ETHICON ENDO-SURGERY, LLC. HARMONIC ACE PLUS 7 ADV HEMOSTASIS 36; INSTRUMENT, ULTRASONIC SURGICAL Back to Search Results
Model Number HARH36
Device Problem Packaging Problem (3007)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/03/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # t9538t.Investigation summary: the analysis results found that the harh36 device was returned inside its package unopened.Upon visual inspection, it was observed that the blister from the packaging was damaged; it was noted to be broken and still adhered to the tyvek.The manufacturing records were reviewed and the manufacturing/packaging criteria were met prior to the release of this batch.Due to the damages found on the blister, a possible cause for this condition is due to improper handling during transit or storage; it appears that the package hit a hard surface and this caused the reported event.The reported complaint was confirmed.All ees product is 100% inspected prior to release.The information you provided is compiled monitored and reviewed on a routine basis for any associated trends.
 
Event Description
It was reported that when the harh36 was removed from the white cardboard shipping container, it was observed that the clear plastic container of the device was cracked on the corner.There were no patient consequences reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HARMONIC ACE PLUS 7 ADV HEMOSTASIS 36
Type of Device
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kara ditty-bovard
475 calle c
guaynabo 00969
*   00969
6107428552
MDR Report Key9841806
MDR Text Key200609138
Report Number3005075853-2020-01652
Device Sequence Number1
Product Code LFL
UDI-Device Identifier10705036014621
UDI-Public10705036014621
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K132612
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 02/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberHARH36
Device Catalogue NumberHARH36
Device Lot NumberT9538T
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/21/2020
Initial Date Manufacturer Received 02/25/2020
Initial Date FDA Received03/17/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/26/2019
Is the Device Single Use? Yes
Type of Device Usage N
Patient Sequence Number1
-
-