(b)(4).Batch # t9538t.Investigation summary: the analysis results found that the harh36 device was returned inside its package unopened.Upon visual inspection, it was observed that the blister from the packaging was damaged; it was noted to be broken and still adhered to the tyvek.The manufacturing records were reviewed and the manufacturing/packaging criteria were met prior to the release of this batch.Due to the damages found on the blister, a possible cause for this condition is due to improper handling during transit or storage; it appears that the package hit a hard surface and this caused the reported event.The reported complaint was confirmed.All ees product is 100% inspected prior to release.The information you provided is compiled monitored and reviewed on a routine basis for any associated trends.
|