Catalog Number 0998-00-0800-53
Device Problem Temperature Problem (3022)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/25/2020
Event Type  malfunction  
Manufacturer Narrative
The production device history record (dhr) for this intra-aortic balloon pump (iabp) was not required to be reviewed as per company standard operating procedure since the device manufacture date is greater than one year from the event date.A getinge service territory manager (stm) was dispatched to evaluate the iabp unit.The stm was unable to reproduce the overheating fault but found fault 77 -compressor motor speed adj - assisting" in the logs.The stm ran the recommended "drive regulator calibration" tests.The vacuum passed the pressure set point test.The stm will order a new scroll compressor assembly, as recommended in the manual, and return to replace it.Upon receipt, the stm returned to replace the scroll compressor which alleviated the fault 77s but the iabp then failed the drive manifold test.The stm replace the pressure and vacuum regulators and recalibrated.The stm completed all safety, functionality and calibration checks and all tests passed to factory specifications.The iabp unit was cleared for clinical use and released to the customer.
 
Event Description
It was reported that the cardiosave intra-aortic balloon pump (iabp) was overheating.It is unknown under which circumstances the event occurred or if a patient was involved; however there was no adverse event reported.
 
Event Description
It was reported that the cardiosave intra-aortic balloon pump (iabp) was overheating.It is unknown under which circumstances the event occurred or if a patient was involved; however there was no adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
MDR Report Key9842792
MDR Text Key192400250
Report Number2249723-2020-00433
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 05/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number0998-00-0800-53
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 02/25/2020
Initial Date FDA Received03/17/2020
Supplement Dates Manufacturer Received04/30/2020
Supplement Dates FDA Received05/01/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-