• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ATRICURE, INC. ATRICLIP LAA EXCLUSION SYSTEM WITH PRELOADED GILLINOV COSGROVE CLIP; ATRICLIP® LAA EXCLUSION SYSTEM WITH PRELOADED GILLINOV COSGROVE¿ CLIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ATRICURE, INC. ATRICLIP LAA EXCLUSION SYSTEM WITH PRELOADED GILLINOV COSGROVE CLIP; ATRICLIP® LAA EXCLUSION SYSTEM WITH PRELOADED GILLINOV COSGROVE¿ CLIP Back to Search Results
Model Number PRO250
Device Problem Use of Device Problem (1670)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/19/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The atriclip pro250 device lot number 96782 was returned for evaluation and was visually and functionally tested.The complaint was confirmed for articulation movement failure and difficulty in deployment.
 
Event Description
It was reported that on (b)(6) 2020 during an off-pump thoracoscopic ablation with left atrial appendage management (laam) procedure, the surgeon placed a pro250 and confirmed clip placement by transesophageal echocardiogram (tee).When the surgeon went to deploy the clip by pulling the orange tab, it seemed jammed and the clip would not release from the handle.At that point, the surgeon decided to use an endoscopic scissor to cut and release the clip.There was no reported procedural complication or adverse consequence to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATRICLIP LAA EXCLUSION SYSTEM WITH PRELOADED GILLINOV COSGROVE CLIP
Type of Device
ATRICLIP® LAA EXCLUSION SYSTEM WITH PRELOADED GILLINOV COSGROVE¿ CLIP
Manufacturer (Section D)
ATRICURE, INC.
7555 innovation way
mason OH 45040
Manufacturer (Section G)
ATRICURE, INC.
7555 innovation way
mason OH 45040
Manufacturer Contact
john ehlert
7555 innovation way
mason, OH 45040
5136448220
MDR Report Key9844886
MDR Text Key195177311
Report Number3011706110-2020-00010
Device Sequence Number1
Product Code FZP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K160454
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 03/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/01/2022
Device Model NumberPRO250
Device Catalogue NumberA000968
Device Lot Number96782
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/26/2020
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/19/2020
Initial Date FDA Received03/17/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/12/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-