• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. LIGAMAX-5MM ENDO CLIP APPLIER; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. LIGAMAX-5MM ENDO CLIP APPLIER; CLIP, IMPLANTABLE Back to Search Results
Model Number EL5ML
Device Problem Misfire (2532)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/12/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch #: t94g01.Investigation summary: the analysis results found that the el5ml device was received inserted into a b5lt trocar with one jaw disengaged from the cam.This condition would not allow the jaws to collapse in order to pull out the el5ml device from the trocar.In order to evaluate the performance of the device, the jaw was readjusted and in the next activation, 10 conforming clips were fed and formed.Finally, the device locked out as intended.Possible causes for the condition of the jaw disengaging from the cam may be inadvertent force, twisting or pressure being placed on the device jaws, using the jaws of the device as a dissector/retractor, or damage to the jaws while entering the trocar.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.Attempts were being made to obtain the following information: what is meant by the device misfired? did the device fire clips and then stop firing? did the clips drop or eject from the jaws of the device? did the device not fire at all? were the clips malformed, scissored? to date, no response has been provided.If the device or further details are received at a later date, a supplemental medwatch will be sent.
 
Event Description
It was reported that during a laparoscopic cholecystectomy, the device misfired.A similar device was used to complete the case.There were no patient consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGAMAX-5MM ENDO CLIP APPLIER
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kara ditty-bovard
475 calle c
guaynabo 00969
*   00969
6107428552
MDR Report Key9847171
MDR Text Key199084047
Report Number3005075853-2020-01679
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705036001843
UDI-Public10705036001843
Combination Product (y/n)N
PMA/PMN Number
K050344
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEL5ML
Device Catalogue NumberEL5ML
Device Lot NumberT94G01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/21/2020
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/21/2020
Initial Date FDA Received03/18/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/20/2019
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-