(b)(4).Batch #: t94g01.Investigation summary: the analysis results found that the el5ml device was received inserted into a b5lt trocar with one jaw disengaged from the cam.This condition would not allow the jaws to collapse in order to pull out the el5ml device from the trocar.In order to evaluate the performance of the device, the jaw was readjusted and in the next activation, 10 conforming clips were fed and formed.Finally, the device locked out as intended.Possible causes for the condition of the jaw disengaging from the cam may be inadvertent force, twisting or pressure being placed on the device jaws, using the jaws of the device as a dissector/retractor, or damage to the jaws while entering the trocar.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.Attempts were being made to obtain the following information: what is meant by the device misfired? did the device fire clips and then stop firing? did the clips drop or eject from the jaws of the device? did the device not fire at all? were the clips malformed, scissored? to date, no response has been provided.If the device or further details are received at a later date, a supplemental medwatch will be sent.
|