Model Number 97000HS |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/18/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation conclusion: triage labeling does not make any claims to correlate to the stratus method for ckmb.Various ckmb assays are not standardized and therefore may have a difference in values depending on the assay used.The customers complaint was not replicated with in-house testing of retain lot t10774n.No issues with ckmb recovery were observed.Manufacturing batch records for lot t10774n were reviewed and found that the lot met final release specifications.The root cause could not be determined.Based on the information available, there is no indication of a product deficiency and no corrective action is required.
|
|
Event Description
|
Customer reported poor correlation for ckmb between triage platform and stratus cs.Customer stated they have not used the triage for any patient reporting.The correlations were done anonymously, therefore no patient details or information can be obtained.Triage ckmb: 3.6 compared to stratus ckmb: 8.4.Triage ckmb: 4.6 compared to stratus ckmb: 2.8.Sites ckmb reference range for the platforms: triage <4.3.Stratus 0.0-5.0.
|
|