• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION MAXZERO NEEDLELESS CONNECTOR; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION MAXZERO NEEDLELESS CONNECTOR; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number MZ1000
Device Problem Fluid/Blood Leak (1250)
Patient Problem Hypoglycemia (1912)
Event Type  Injury  
Manufacturer Narrative
Although requested, device has not been received.A follow up report will be submitted with failure investigation results should the device be received for evaluation.
 
Event Description
The reported feedback suggests that there was a leakage of drug from the device.From the reported information, the patient suffered hypoglycemia as a result of this incident.
 
Manufacturer Narrative
One mz1000 sample from lot 19115073 was received in sealed packaging and one empty packaging from lot 19106164 was received for investigation of complaint reference (b)(4).In addition, one alargador extension set and one heuer three-way tap was also returned to assist the investigation.A visual inspection of the returned mz1000 sample did not identify any product defects or manufacturing issues which could have caused or contributed to the customer¿s experience.In an attempt to replicate the customer's experience, the returned extension set and three-way tap were connected to the maxzero component; in each instance no leakage was identified.Pressure test also did not identify any leakage from the mz1000 sample.The details of this feedback were forwarded to the manufacturing site for investigation.In this instance, a definitive root cause could not be identified as testing of the returned samples did not identify any product defects that could have contributed to the customer¿s experience.A review of the production records for lot 19115073 and 19106164 did not identify any in-process testing failures or quality deviations which may have resulted in a report of this nature.
 
Event Description
The reported feedback suggests that there was a leakage of drug from the device.From the reported information, the patient suffered hypoglycemia as a result of this incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAXZERO NEEDLELESS CONNECTOR
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
CAREFUSION
10020 pacific mesa blvd
san diego CA 92121 4386
MDR Report Key9857917
MDR Text Key184287196
Report Number9616066-2020-00829
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
PMA/PMN Number
K132413
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMZ1000
Device Catalogue NumberMZ1000
Device Lot Number19106164
Was Device Available for Evaluation? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received03/19/2020
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received04/08/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-