It was reported that customer saw white crystallization coming out of the fill port as well as the cartridge tubing.Customer experienced moderate levels of ketones and an elevated blood glucose (bg) level of over 600mg/dl and was subsequently hospitalized.Reportedly, the ketone level was not considered dangerous by the healthcare provider.In an attempt to resolve elevated bg, the customer administered a bolus, manual injections, and tested for ketones.While at the hospital, the customer was treated with insulin drip, and fluids.The customer was released from the hospital approximately two days later with no permanent damage and issue resolved.The customer alleged that the insulin crystallization was the cause of elevated bg.Reportedly, customer had been using the same insulin, cartridge, and infusion sets for a week.Tandem technical support advised the customer to only use pump supplies for up to 3 days as using for a longer time is considered off label.
|