Model Number N/A
Device Problem Difficult to Insert (1316)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/02/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Concomitant medical products: 42532007102, tibia cemented 5 degree stemmed right size e, 64452029.Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the product was discarded by the hospital.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that during an initial knee arthroplasty, the poly would not seat.A different poly was used and successfully implanted.There were no consequences or impact to the patient.
 
Manufacturer Narrative
No product was returned or pictures provided; visual and dimensional evaluations could not be performed.The dhr was reviewed and no discrepancies relevant to the reported event were found.The root cause of the reported event is attributed to user error.As stated on the label for part 42522100910, the articular surface is to be used with a size g or h tibia.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTICULAR SURFACE MEDIAL CONGRUENT (MC) RIGHT 10 MM HEIGHT
Type of Device
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI 
MDR Report Key9867931
MDR Text Key191580150
Report Number3007963827-2020-00092
Device Sequence Number1
Product Code MBH
UDI-Device Identifier00889024468498
UDI-Public00889024468498
Combination Product (y/n)N
PMA/PMN Number
K150090
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Type of Report Initial,Followup
Report Date 04/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number42522100910
Device Lot Number64453825
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/02/2020
Initial Date FDA Received03/23/2020
Supplement Dates Manufacturer Received04/28/2020
Supplement Dates FDA Received04/29/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-