• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD MEDICAL (BD WEST) MEDICAL SURGICAL 10 ML BD POSIFLUSH¿ SP PRE-FILLED FLUSH SYRINGE NACL 0.9%; N/A

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD MEDICAL (BD WEST) MEDICAL SURGICAL 10 ML BD POSIFLUSH¿ SP PRE-FILLED FLUSH SYRINGE NACL 0.9%; N/A Back to Search Results
Catalog Number 306575
Device Problem Leak/Splash (1354)
Patient Problem No Patient Involvement (2645)
Event Date 03/10/2020
Event Type  malfunction  
Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.Investigation summary: one sample was received.It has no packaging flow wrap.It has the plunger rod-rubber stopper, the tip cap and saline solution.It has the luer tip bent.This occurs when the syringe is misplaced in the sterilizer tray the next tray that goes on top lays on top of the misplaced syringe inducing the damage of the luer tip.A device history record review was completed with zero defects found.No quality notifications were written for this batch, nor for the associated assembly batches.Investigation conclusion: complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.Root cause description: it occurred during the sterilizer tray loading process.Rationale: capa not required at this time.
 
Event Description
It was reported that 10 ml bd posiflush¿ sp pre-filled flush syringe nacl 0.9% tip was bent.This was discovered before use.The following information was provided by the initial reporter: bent tip.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
10 ML BD POSIFLUSH¿ SP PRE-FILLED FLUSH SYRINGE NACL 0.9%
Type of Device
N/A
Manufacturer (Section D)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer (Section G)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer Contact
brett wilko
9450 south state street
sandy, UT 84070
8015652341
MDR Report Key9875639
MDR Text Key187055656
Report Number1911916-2020-00295
Device Sequence Number1
Product Code NGT
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date07/31/2022
Device Catalogue Number306575
Device Lot Number9212795
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/20/2020
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/10/2020
Initial Date FDA Received03/24/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/31/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-