Model Number 1456Q/86
Device Problem Low impedance (2285)
Patient Problem Undesired Nerve Stimulation (1980)
Event Date 03/11/2020
Event Type  malfunction  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that when the patient presented for magnetic resonance imaging (mri), the device was checked before mri.The patient was dependent and severe phrenic nerve stimulation was noted on the left ventricular lead.While programming the device to mri settings, the left ventricular (lv) lead impedance exhibited lower out of range value according to the device settings.The mri procedure aborted, and decision was taken to monitor the lead for any possible malfunction.Programming changes were discussed to make the system mri conditional.The patient was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET LEAD SMALL-S, 86 CM
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
elizabeth boltz
15900 valley view court
sylmar, CA 91342
MDR Report Key9888066
MDR Text Key185198421
Report Number2938836-2020-02246
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734510158
UDI-Public05414734510158
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 03/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2022
Device Model Number1456Q/86
Device Catalogue Number1456Q/86
Device Lot NumberA000085058
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/12/2020
Initial Date FDA Received03/27/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/21/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age39 YR
-
-