Model Number 21AGFN-756
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/18/2019
Event Type  malfunction  
Manufacturer Narrative
An event of a dislodged leaflet was reported.The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications at the time of commercialization.Based on the information received, the cause of the reported incident could not be conclusively determined.Please note, per the instructions for use artmt100122074 ver.A, "use only the valve holder/rotator packaged with the sjm regent¿ mechanical heart valve to perform valve rotation.Use of other instruments could result in structural damage.The valve holder/rotator is intended for single use only and should be discarded after surgery." "to avoid structural damage, the valve must be rotated in the fully closed position." "to minimize rotational torque, verify that the valve holder/rotator is properly seated in the valve, and that the valve holder handle is perpendicular to the valve.".
 
Event Description
On (b)(6) 2019, a 21mm regent valve was selected to be implant.During the procedure, one of the leaflet dislodged to the left ventricle.The valve was removed from the patient and replaced with another 21mm regent valve was implanted.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM REGENT HEART VALVE W/FLEX CUFF
Type of Device
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ST. JUDE MEDICAL PUERTO RICO, INC.
p.o. box 998
lot 20 b st.
caguas, puerto rico 00725
RQ  00725
Manufacturer (Section G)
ST. JUDE MEDICAL PUERTO RICO, INC.
p.o. box 998
lot 20 b st.
caguas, puerto rico 00725
RQ   00725
Manufacturer Contact
pamela yip
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key9896103
MDR Text Key194332083
Report Number2648612-2020-00039
Device Sequence Number1
Product Code LWQ
UDI-Device Identifier05414734005975
UDI-Public05414734005975
Combination Product (y/n)N
Reporter Country CodeEG
PMA/PMN Number
P810002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Remedial Action Replace
Type of Report Initial
Report Date 03/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/12/2023
Device Model Number21AGFN-756
Device Catalogue Number21AGFN-756
Device Lot Number6587165
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/02/2020
Initial Date FDA Received03/29/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/13/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-