MICROMAR INDUSTRIA E COMERCIO LTDA EASYDRILL PERFORATOR; DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOUND, POWERED)
|
Back to Search Results |
|
Model Number DM0010FAA |
Device Problem
Device Remains Activated (1525)
|
Patient Problems
Perforation (2001); Injury (2348)
|
Event Date 03/02/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
On follow up, it was confirmed that there was a dural tear as well as a damage to the cortex.It was also mentioned that there was a delay in the procedure but it was completed using a back up device.Additional information provided that there was a non-routine action taken because of the event.Further information about the event is being followed up from the facility contact person.
|
|
Manufacturer Narrative
|
No conclusion can be drawn.No evaluation was performed, as the devices were not returned.If the devices are returned in the future, product analysis may be performed.This dm0010faa easydrill cranial perforator with lot number 1600/19 was manufactured by micromar.Multiple warnings are included in the easydrill cranial perforator ifu manual including: it is essential to keep the drill perpendicular (90°) at the predetermined point of the skull to be drilled, as an excessive deviation from perpendicularity may cause the product to fail and lead to serious patient injury.Select a drill bit suitable for the bone thickness.This prevents the drill from tearing the bone or brain tissue (similar effect when the bit is not positioned at 90°).If the components of the easydrill cranial perforator become loose during trephination, discontinue use.The drill can cut or tear the dura mater when it unlocks.Check some conditions before performing the procedure, such as the existence of adhering dura mater or other anomalies adjacent to the drilling site.We will continue to track and trend this complaint type.(b)(6).If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that during a temporal tumor resection, the perforator clutch did not disengage.It was reported that there was patient impact.Additional information about the event is being followed up from the facility contact person.
|
|
Event Description
|
On further follow up, it was reported that there was a 5-minute procedure delay due to continued hemostasis and dissection for addit ional blood control.It was also confirmed that the patient had no further issues post-surgery.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|