After review of the medical records patient progress note dated (b)(6) 2018, reported of clicking about her left hip and significantly elevated metal levels.X-ray shows a fracture at the base of the clothespin part of the stem and loosening of the screw.Patient took a fall and since then experienced clicking.Both hips were metal on metal articulation.There was no left hip revision reported.Doi: (b)(6) 2009.Dor: none reported, left hip.Please see (b)(4) for the right hip revision.
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.
|