It was reported that an emboshield nav6 embolic protection system (eps) was opened and it was noted that the deliver catheter (dc) pod was wrinkled and the filter could not be inserted into the dc pod.The eps was not used.A second emboshield nav6 eps was being prepared; however, the filter could not be inserted into the dc pod.The eps was not used.The procedure was successfully completed with a new emboshield nav6 eps.There were no adverse patient effects and no clinically significant delay in the procedure.The return device analysis identified that both eps's showed signs of use.The eps's were returned with blood on the coils and contrast on the dc pods, shafts, and coils.The return device analysis for the first eps identified that the proximal coil was separated (not in two separate pieces) 11.5mm, 12mm and 5.5cm distal to the proximal solder; however, the core was still intact.The return device analysis for the second eps identified that the dc pod was separated approximately 2.5mm distal to the marker.There was a tear on the distal separated portion of the dc pod for a length of 3.5mm.No additional information was provided.
|