• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY GMBH HEART LUNG MACHINE; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY GMBH HEART LUNG MACHINE; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 701027652 - HL 20
Device Problem Pumping Problem (3016)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/20/2020
Event Type  malfunction  
Manufacturer Narrative
It was reported that the double pump displayed the error message head error.No service order was provided.However according to the failure description the optical tacho board (part number 70100.7815) was exchanged and the pump worked normally.A similar complaint (b)(4) was already investigated on 2019-08-23.According to the life cycle engineering report# lce3993 dated in 2019-08-26 following was performed: the failure was reproducible and the most probable root cause is a broken wiring of the tachoboard.The reported failure "head error" could be confirmed.It is unknown when the event occurred, was the device being used for treatment or diagnosis of the patient.The hl20 in question was responsible for this complaint/event.The occurrence rate was calculated for the reported issue and it was determined that this is not a systemic issue.Therefore, no remedial action is required the occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopulmonary¿s trending program and additional investigations or corrections will be implemented in case of adverse trending.
 
Event Description
It was reported that the double pump displayed the error message head error (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEART LUNG MACHINE
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
Manufacturer (Section G)
NURSEL BOELENS
maquet cardiopulmonary ag
kehler strasse 31
rastatt 76437
GM   76437
Manufacturer Contact
maquet cardiopulmonary ag
kehler strasse 31
rastatt 76437
4972229321
MDR Report Key9900961
MDR Text Key199067971
Report Number8010762-2020-00121
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
PMA/PMN Number
K943803
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign,health profe
Type of Report Initial
Report Date 03/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/30/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number701027652 - HL 20
Device Catalogue Number701027652
Date Manufacturer Received03/23/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-