Model Number 16414
Device Problem No Device Output (1435)
Patient Problem No Patient Involvement (2645)
Event Date 03/06/2020
Event Type  malfunction  
Manufacturer Narrative
The fsr installed a new cpg temperature/pressure printer circuit board (pcb) and tested.The unit operated to the manufacturer's specifications.The suspect device was returned to the manufacturer for further evaluation.
 
Event Description
The field service representative (fsr) reported that during preventive maintenance (pm) of the device, neither of the three temperature ports, cardioplegia (cpg), a, or b would work on cpg monitor.None of the temperature values were being displayed.There was no patient involvement.
 
Manufacturer Narrative
The reported complaint was confirmed.During laboratory analysis, the product surveillance technician observed there to be no output from the printed circuit board (pcb).When he installed the pcb into a lab use only cardioplegia (cpg)monitor, the temperature display was blank.A working cpg monitor will display dashes if there is not a temperature probe connected.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SARNS 8000 CARDIOPLEGIA MONITOR
Type of Device
COMPUTER, BLOOD-PRESSURE-SARNS 8000 CARDIOPLEGIA MONITOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key9901051
MDR Text Key199544386
Report Number1828100-2020-00144
Device Sequence Number1
Product Code DSK
Combination Product (y/n)N
PMA/PMN Number
K960916
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 06/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number16414
Device Catalogue Number16414
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/09/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/06/2020
Initial Date FDA Received03/30/2020
Supplement Dates Manufacturer Received05/20/2020
Supplement Dates FDA Received06/01/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-